The obvious answer to where platinum rhodium wire sits in a thermocouple is that it sits at the measuring end, and that answer is incomplete in a way that costs money. The wire occupies the hot segment of the circuit only, from the junction to the connection head. Everything after the head is a different cable, and for the platinum types that cable is not made of platinum at all.
![]()
A temperature reading is produced by the whole path between the process and the instrument. Each stage introduces its own error, and each stage is bought and specified separately.
| Stage | What it is | What it is made of | Where its error comes from |
|---|---|---|---|
| The hot junction | The point where the two legs are welded together and touch the process | The two thermocouple alloys | The wire tolerance, and how well the junction is placed |
| The thermocouple wire | The run from the junction to the connection head | Platinum rhodium and platinum for Types R and S; two platinum rhodium alloys for Type B | The wire tolerance, and contamination at temperature |
| The connection head | The terminal block where the sensor ends and the cable begins | A terminal block, which has to hold both terminals at the same temperature | A temperature difference across the block, which is added to the reading |
| The run to the instrument | The long cable from the head to the panel | Extension cable or compensating cable, depending on the type | The cable tolerance, and the temperature along the route |
| The reference junction and instrument | The terminals at the instrument, and the electronics behind them | Copper, and the instrument input | The accuracy of the reference junction measurement and the instrument |
The wire described in the title is stage two. It is the only stage made of platinum, and it is usually the shortest part of the circuit. A furnace sensor may have 300 mm of platinum rhodium wire and 30 metres of something else behind it.
If the whole circuit had to be the thermocouple alloy, the cost of a platinum installation would follow the cable length, and a long run would be unaffordable. The standards solve this by allowing a second kind of cable, and the distinction between the two is marked in the cable code itself.
| Mark | Meaning | Conductors | Where it is used |
|---|---|---|---|
| X | Extension cable. The conductors are the same alloys as the thermocouple | For example KX is nickel chromium against nickel aluminium, the same pair as the Type K wire | Base metal types, where the metal is cheap enough to run the whole way |
| C | Compensating cable. The conductors are substitute alloys chosen to match the thermocouple's voltage over a limited range | For example KCA uses iron against constantan, which is not what a Type K junction is made of | Noble metal types, and any base metal run that stays in a controlled ambient temperature |
Types R, S and B have no true extension cable in common use, and the reason is the metal itself. A platinum rhodium compensating cable is not made of platinum rhodium, because nobody would pay to run those alloys through a cable tray. What is used instead is a copper-based pair whose output matches the platinum pair closely enough over a limited band of ambient temperature.
Published cable data shows how the R and S compensating conductors have converged. The older RCA and SCA pairs, copper against a low-value copper nickel, have largely been replaced by a single combination, RCB and SCB, which uses copper against a copper nickel molybdenum alloy and serves both Type R and Type S. One alloy now covers the two types, which is worth knowing when an old drawing names RCA or SCA.
A cable tolerance is not the size of the error the cable is allowed to have. It is the extra error the cable may add to a measurement that was already correct, and it is quoted against a temperature band for the conductors.
| Cable | Conductors | IEC Class 1 | IEC Class 2 | ANSI Class 2 | Cable temperature |
|---|---|---|---|---|---|
| KX | Nickel chromium / nickel aluminium | ±1.5 °C | ±2.5 °C | ±2.2 °C | −25 to 200 °C |
| JX | Iron / constantan | ±1.5 °C | ±2.5 °C | ±2.2 °C | −25 to 200 °C |
| TX | Copper / constantan | ±0.5 °C | ±1.0 °C | ±1.0 °C | −25 to 200 °C |
| EX | Nickel chromium / constantan | ±1.5 °C | ±2.5 °C | ±1.7 °C | −25 to 200 °C |
| NX | Nicrosil / nisil | ±1.5 °C | ±2.5 °C | not specified | −25 to 200 °C |
| KCA | Iron / constantan | not specified | ±2.5 °C | ±2.2 °C | 0 to 150 °C |
| KCB | Copper / constantan | not specified | ±2.5 °C | ±2.2 °C | 0 to 100 °C |
| NC | Copper nickel magnesium both legs | not specified | ±2.5 °C | ±2.2 °C | 0 to 150 °C |
| RCB and SCB | Copper / copper nickel molybdenum | not specified | ±5.0 °C | ±5.0 °C | 0 to 200 °C |
Two numbers in that table deserve attention. The first is that the R and S compensating combination carries a Class 2 tolerance of ±5.0 °C, three times the figure for a KX extension cable. The second is that the tolerances are quoted for the conductors, and the published note adds that the usable range may be narrower still because of the insulation, which is a question for the cable maker, not for the standard.
The consequence is uncomfortable for a platinum installation. A special-limits Type S wire is held to ±0.6 °C or 0.1 % of reading, and the compensating cable behind it can add five degrees. Buying the tightest wire and the cheapest cable relocates the error instead of removing it, which is why the cable belongs in the accuracy calculation and not in the accessories column.
How much the cable matters is set by how large the signal is to begin with. Published values for the voltage produced at 100 °C show the gap between the base metal types and the platinum ones.
| Type | Voltage at 100 °C |
|---|---|
| E | 6.32 mV |
| J | 5.27 mV |
| T | 4.28 mV |
| K | 4.10 mV |
| N | 2.77 mV |
| R and S | 0.65 mV |
| B | About 0.03 mV, which is why published compensating data for Type B is quoted at higher reference temperatures instead |
A Type R or Type S pair generates about one sixth of the voltage a Type K pair does at the same temperature. The instrument has to resolve a smaller signal, and every avoidable junction in the circuit contributes a proportionally larger share of the total. That is the practical reason a platinum installation is more sensitive to how it is wired than a base metal one, and it has nothing to do with the wire itself.
| Point | What it requires | What goes wrong otherwise |
|---|---|---|
| The connection head | Both terminals held at the same temperature, on an isothermal block in the head | A gradient across the block becomes a voltage that is added to the reading, and it looks like a wire fault |
| The polarity of every pair | Positive leg to positive terminal at every joint, including the cable joints | A reversed pair inverts the signal, and the result reads like a calibration problem |
| Ordinary copper anywhere in the run | Not used. Copper introduces junctions between unlike metals that were never in the calculation | Parasitic voltages appear at each copper joint, and they move with the ambient temperature |
| The ambient temperature along the route | Kept inside the cable temperature range, and stable | A compensating cable taken through a hot area is being asked to work outside the band it was matched over |
| Loop resistance | Kept low, with below about 100 ohms as the usual working guide | High loop resistance with a millivolt signal invites noise and loss, and some instruments refuse the input |
| Screening | A shield and a drain, because the signal is in millivolts | Electrical noise from drives and heaters is read as temperature |
| Identification | The cable code and the sheath colour checked before the core is stripped | IEC colours put the negative leg in white, and a sheath that follows the positive leg is blue instead on intrinsically safe circuits, which is the one case where the rule changes |
The intrinsically safe exception is worth isolating, because it inverts the usual rule. Under IEC 60584-3 the sheath normally takes the colour of the positive leg, and on an intrinsically safe circuit the sheath is blue regardless of type. A blue sheath is therefore information about the circuit and not about the thermocouple.
| Item | Why it belongs on the order |
|---|---|
| The wire: type, grade or class, and diameter | Sets the signal the instrument expects and the error allowed at the junction |
| The length of the wire segment | The platinum wire is bought by length, and the run it covers is the short one |
| The cable: code, such as RCB, and its tolerance class | The cable code states which conductors are inside, and the class states the error they may add |
| The ambient temperature along the cable run | The compensating pairs are matched over a band, and the band for R and S is 0 to 200 °C for the conductors |
| The insulation and the sheath, including whether the circuit is intrinsically safe | The insulation may restrict the usable range below the conductor figure, and the sheath colour convention changes on an intrinsically safe circuit |
| Screening and drain | The signal is small enough that noise is a real contribution to the total error |
In the hot segment, from the measuring junction to the connection head, and nowhere else on a normal installation. The wire is the part that has to survive the process temperature, so it is kept as short as the geometry allows. The run from the head to the instrument is a separate cable, and for Types R, S and B that cable is not platinum.
The conductors. An extension cable uses the same alloys as the thermocouple, so the circuit behaves as though the sensor were longer. A compensating cable uses substitute alloys selected to reproduce the same voltage over a limited temperature band, which makes it far cheaper and limits how hot the route may be. The cable code says which one is inside: X for extension, C for compensating.
Because it would be made of platinum and rhodium, and the cost would follow the cable length. A compensating pair is used instead. Published data shows that the older RCA and SCA combinations have largely been replaced by RCB and SCB, a copper against copper nickel molybdenum pair that serves both types with a Class 2 tolerance of ±5.0 °C over 0 to 200 °C.
It can, and on the platinum types it often does. A Type R or Type S pair produces about 0.65 mV at 100 °C, roughly a sixth of what a Type K pair produces, so the same avoidable error is a larger fraction of the signal. The published Class 2 tolerance for the R and S compensating cable is ±5.0 °C, against ±1.5 °C for a KX extension cable, so the cable can contribute more error than the wire is held to.
No. Every joint between unlike metals produces its own voltage, so copper in the middle of a thermocouple circuit adds junctions that were never part of the measurement, and those voltages move with the ambient temperature. The cable has to match the type from the head to the instrument, and the joints have to be made with the correct polarity.
The cable type, with one exception that catches people out. Under IEC 60584-3 the negative leg is white and the sheath follows the colour of the positive leg. On an intrinsically safe circuit the sheath is blue whatever the type, so a blue sheath is telling you about the circuit, not about the thermocouple inside it.
Platinum rhodium wire sits at the measuring end, and that is the whole of its extent: the junction, and the short run to the connection head. The long part of the circuit is extension or compensating cable, and the noble metal types are served by the compensating kind because a platinum cable would be priced by the metre. The cable is not a neutral accessory. Its tolerance is added to the wire tolerance, the signal it carries is about a sixth the size of a base metal one, and the loop resistance, the screening and the polarity all decide whether the reading arrives intact.
DLX Alloy has made nickel and nickel alloy products at Changzhou, Jiangsu since 2002, on a 12,000 m² site certified to ISO 9001 and audited by SGS. Platinum rhodium thermocouple wire is supplied in the B, R and S types, and in the bare wire form the temperature loop needs: send the type, the diameter, the grade and the loop length for a confirmed quotation.
The figures quoted are published values and are not acceptance limits. A cable tolerance is the additional error a cable may introduce under the conditions the standard specifies, and it is not a statement about a complete installation. What the loop achieves depends on the junction, the head, the cable, the reference junction and the instrument together, and that is a commissioning question, not a catalogue one.